Description

The aim of this work was to study the applicability of silicon microdosimeters for high LET ion measurements. The response and radiation damage of two silicon on insulator (SOI) 3D microdosimeters developed by the Centre for Medical Radiation Physics (CMRP) were investigated with a range of different low energy ions, with high linear energy transfer (LET). The two microdosimeters n-SOI and p-SOI were irradiated with a number of different ions including $^{7}$Li, $^{12}$C, $^{16}$O and $^{48}$Ti with ranges below 350 µm in silicon.

The two detectors used in this work are called the Bridge and the Mushroom. The Bridge microdosimeter is based on an array of planar 30 x 30 x 10 µm cubic SVs fabricated on a high resistivity of 3 kΩ.cm n-SOI active layer of thickness 10 µm and low resistivity supporting wafer [1]. The Mushroom microdosimeter structure used in this work is called a trenched 3D and it consists of 3D cylindrical SVs with a core columnar n+ region and each SV is surrounded with p+ trench to form a p-n junction. The Mushroom microdosimeter has a thickness of 9.1 um and diameter of 30 µm fabricated on high resistivity p-type silicon (> 10 kΩ.cm).

Using a specially generated low intensity beam (beam fluence of approximately 1200 particles/s), irradiations were conducted at The Heavy Ion Accelerator Facility (HIAF) at the Australian National University using both the Bridge and Mushroom 3D microdosimeters. Radiation damage of the Bridge and Mushroom microdosimeters was studied using the ion beam induced charge collection technique (IBIC) at the 6 MV accelerator SIRIUS, located at the Centre for Accelerator Science (CAS) facility at ANSTO. This system includes a Confocal Heavy Ion Micro-Probe (CHIMP) which is capable of delivering Carbon, Helium and Hydrogen ions with energies of 24 MeV, 5.5 MeV and 8 MeV, respectively.

Results presented will show that no plasma effects were seen in the SOI microdosimeters when irradiated with the high LET ions. A Monte Carlo simulation using Geant4 was compared to the experimental measurements, whereby some discrepancies were observed for heavier ions at lower energies. This discrepancy can be partly attributed to uncertainties in the thickness of the energy degraders and overlayers of the devices. The radiation hardness of the two devices was studied using the Ion Beam Induced Charge Collection technique (IBIC). Charge buildup was seen outside of the SV in the SiO$_2$ layer after irradiation, however both types of microdosimeters when biased had no essential changes in charge collection efficiency (CCE) in the SV after irradiation with low energy ions. IBIC results for both detector types will be presented in full

References:
[1] L. T. Tran, L. Chartier, D. Bolst, et al., “3D silicon microdosimetry and rbe study using $^{12}$C ion of different energies,” IEEE Trans Nucl Sci, vol. 62, no. 6, pp. 3027–3033, Dec 2015.

Do yo wish to take part in the poster slam No
Travel Funding No
Level of Expertise Student
Speakers Gender Male

Primary author

Benjamin James (UOW - CMRP)

Co-authors

Linh Tran (Centre for Medical Radiation Physics, University of Wollongong) Mr James Vohradsky (UOW) Mr David Bolst Vladimir Pan (University of Wollongong) Ms Madeline Carr Mrs Susanna Guatelli Mr Alex Pogossov Mr Marco Petasecca Michael Lerch (University of Wollongong) Mr Dale Prokopovich Mr Mark Reinhard Zeljko Pastuovic (Centre for Accelerator Science of ANSTO) Mr Marco Povoli Ms Angela Kok Mr David Hinde Ms Mahananda Dasgupta Mr Andrew Stuchbery Mr Vladimir Perevertaylo Mr Anatoly Rosenfeld

Presentation Materials

There are no materials yet.
Your browser is out of date!

Update your browser to view this website correctly. Update my browser now

×